101
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects
and people.
a Field strengths from fixed transmitters, such as base stations for radio (cel-
lular / cordless) telephones and land mobile radios, amateur radio, AM and
FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF trans-
mitters, and electromagnetic site survey should be considered. If the meas-
ured field strength in the location in which The Fingertip Oximeter is used
exceeds the applicable RF compliance level above, The Fingertip Oximeter
should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as re-orienting or
relocating The Fingertip Oximeter.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be
less than 3V/m.