English
1. Disconnect all power from console.
2. Clean console by wiping with non-abrasive cloth and neutral pH detergent
and deionized, distilled, or purified water after each case.
3. Dry thoroughly with non-abrasive cloth.
Foot Pedals
Caution: Do not immerse or sterilize. Do not allow liquid to enter foot pedal.
Note: Foot pedals are non-sterile capital equipment and shall always be
maintained outside the Operating Room Sterile Field.
1. Disconnect from console.
2. Clean foot pedal by wiping with non-abrasive cloth and neutral pH
detergent and deionized, distilled, or purified water after each case.
3. Dry thoroughly with non-abrasive cloth.
Dissection Tools
Warning: Do not reprocess dissection tools. Dissection tools are
disposable and SINGLE USE ONLY.
Dispose of properly in sharps container.
Note: The clinical processing instructions provided have been validated
by DePuy Synthes Power Tools for preparing a non-sterile medical device;
this instruction is provided in accordance with ISO 17664 and ANSI/AAMI
ST81.
Mechanical/Automated Cleaning
Warnings:
• Manual pre-cleaning must be completed prior to mechanical
cleaning.
• Cannulations must be pre-cleaned with appropriate brushes as
described in pre-cleaning instructions.
• At the point of use the device must be cleaned as soon as possible
after use to prevent drying of blood, tissue, other biological debris
and contaminants on the device.
• National regulations must be observed.
• You must follow standard hospital procedures.
• You must follow manufacturer's recommendations for any
detergents, disinfectants and washing equipment used.
• Do not reprocess dissection tools. Dissection tools are disposable
and SINGLE USE ONLY.
6