• Hold handpiece with distal end pointing down and flush
inside of distal end using a syringe or pipette filled with hot
deionized, distilled or purified water.
Caution: Do not immerse. Do not use high pressure water or air.
Do not allow running water to enter distal end of handpiece or
console connector (end of cable).
5. Inspect
• Visually inspect handpiece. Repeat Steps 2 thru 5 if visible
soil remains.
6. Go to Sterilization section for further instructions.
Attachments
Note: The instructions in this section are specific to the G1
Attachments and MA Bearing Sleeves listed in Table 1 in the
Mechanical/Automated Cleaning section of this document.
1. Preparation
• Assemble all necessary supplies including Attachment
Cleaning Brush (ACB), Small Diameter Cleaning Brush
(SDCB), sponge, lint-free cloth and soft bristle brush.
• Remove attachment and dissection tool from handpiece.
2. Rinse
• Rinse attachment under running cold deionized, distilled or
purified water for a minimum of 2 minutes.
• Manipulate any moving parts such as release sleeves under
running cold deionized, distilled or purified water to loosen or
remove gross soil.
• Use a sponge, soft lint-free cloth or soft bristle brush to assist
in removing gross soil.
Caution: Do not use high pressure water or air. Do not rinse
attachment with saline solution.
3. Clean
• Prepare cleaning solution using alkaline detergent or neutral
pH enzymatic detergent per manufacturer's directions for
correct temperature, water quality (i.e. pH, hardness) and
product concentration/dilution.
• Fully immerse attachment in cleaning solution in a suitable
container and manually agitate for 2 minutes.
3