5. Inspect
• Visually inspect attachment for soil. Repeat Steps 2 thru 5 if
visible soil remains.
6. Go to Mechanical/Automated Cleaning section for further
instructions.
Warning: Manual pre-cleaning must be followed by mechanical/
automated cleaning.
Mechanical/Automated Cleaning
Note: Prior to using G1 Cleaning and Sterilization Basket for first
time, attach connector sleeve to basket by clipping onto internal
support structure (Figure 2). The connector sleeve should remain with
basket for future use.
1. Loading the Basket
• Load devices in G1 Cleaning and Sterilization Basket by
securing them in appropriate brackets, as shown in G1
Cleaning and Sterilization Basket Loading Diagram, with
distal end angled down where applicable.
• Place connector sleeve on connector end of electric
handpiece (Figure 3).
Caution: Connector end of electric handpiece must be sealed with
connector sleeve provided with basket (Figure 3).
Warning: Consult G1 Cleaning and Sterilization Basket Loading
Diagram for validated loading configurations. Angled attachments
must be loaded in the correct orientation as shown in Figure 4.
Figure 2
Figure 4
Figure 3
11