Note: Prior to using G1 Cleaning and Sterilization Basket for first time,
attach connector sleeve to basket by clipping onto internal support
structure (Figure 2). The connector sleeve should remain with basket for
future use.
Warning: Devices must be cleaned prior to sterilization using the validated
cleaning method outlined in these instructions.
1. Loading the Basket
• Load devices in G1 Cleaning and Sterilization Basket by securing
them in appropriate brackets, as shown in G1 Cleaning and
Sterilization Basket Loading Diagram, with distal end angled down
where applicable.
• Remove connector sleeve from handpiece connector (Figure 1).
Replace connector sleeve in its bracket and lay handpiece
connector in bottom of basket.
Warning: Handpiece connector sleeve must be removed from the handpiece
connector prior to sterilization (Figure 1). Consult G1 Cleaning and Sterilization
Basket Loading Diagram for validated loading configurations. Angled
attachments must be loaded in the correct orientation as shown in Figure 2.
Figure 1
2. Packaging
• When wrapping the basket the lid must be used.
• In US double wrap basket or instruments with an FDA cleared
sterilization wrap using appropriate wrapping techniques described
in ANSI/AAMI ST79 (For example: Simultaneous double wrapping or
sequential wrapping).
• Outside US double wrap basket or instruments in accordance with
local procedures, using standard wrap and wrapping techniques
such as those describe in ANSI/AAMI ST79.
Warning: When using a sterilization wrap the basket lid must be used.
Must follow manufacturer's instructions for any wraps used. Do not use
individual sterilization pouches. Only use indicated basket for equipment.