To Calibrate the Solaris Multigas Detector (FIGURE 4-1):
1. Turn ON the instrument and verify that battery is sufficiently
charged.
2. Wait until the Measure Gases page appears.
3. Push and hold the RESET button until CAL ZERO? flashes on the
display (FIGURE 4-2).
4. Push the ON-OFF button to zero the instrument.
Instrument must be in fresh air to perform the zero.
CAL ZERO flashes.
NOTE: To skip the Zero procedure and move directly to the
calibration span procedure, push the RESET button.
If no button is pushed for 30 seconds, the instrument
returns to the Measure mode.
Once the zeros are set, CAL SPAN? flashes (FIGURE 4-3).