Using The Solaris Multigas Detector; Turning On The Solaris Multigas Detector; Last Cal Date - MSA Solaris Manual De Operación

Ocultar thumbs Ver también para Solaris:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Chapter 2,

Using the Solaris Multigas Detector

Turning ON the Solaris Multigas Detector

Press the Power ON button; the instrument displays:
1. A self-test:
All segments display
Audible alarm sounds
Alarm light illuminate
Vibrator activates
Software version displays
Internal diagnostics
"VISUAL OFF" displays if red LEDs disabled
"BACKLITE OFF" displays if backlight disabled
"AUDIBLE OFF" displays if buzzer disabled
"VIBRATE OFF" displays if vibrator disabled.
2. Alarm setpoints:
Low
High
STEL (if activated)
TWA (if activated)
3. Calibration gas (expected calibration gas values)
4. Time and date (if data logging option installed)
5. Last CAL date (if data logging option installed)
6. CAL due date (if activated and if data logging option installed)
7. Instrument warm-up period
8. Fresh Air Setup option.

Last Cal Date

The Solaris Multigas Detector is equipped with a "last successful
calibration date" feature. The date shown is the last date that all
installed sensors were successfully calibrated. "LAST CAL" is displayed
with this date in the following format:
MM:DD:YY
If any of the sensors were not previously calibrated,
"LAST CAL, INVALID" is displayed.
2-1