Instructions for use elements e-motion Dental Motor
2 Safety | 2.1 Infection hazard

2 Safety

The instructions for use are a component of the product and must be read care-
fully prior to use and be accessible at all times.
The device may only be used in accordance with the intended use, any other
type of use is not permitted.

2.1 Infection hazard

Patients, users or third parties could be infected by contaminated medical
devices.
▶ Take suitable personal protective measures.
▶ Follow the instructions for use of the components.
▶ Before initial startup and after each use, reprocess the product and ac-
cessories appropriately.
▶ Carry out the reprocessing as described in the instructions for use. The pro-
cedure has been validated by the manufacturer.
▶ If you deviate from this procedure, it is essential to make sure that the re-
processing is effective.
▶ Reprocess the product and accessories appropriately before disposal.

2.2 Explosion hazard area

Electrical sparks in the product can lead to explosion or fire.
▶ Do not use product in explosion hazard areas.
▶ Do not operate the product in an oxygen-enriched atmosphere.
▶ Do not use the product in the vicinity of flammable gases.

2.3 Technical condition

A damaged device or components could injure patients, users and third parties.
A damaged power cable or missing protective conductor can lead to electrical
shock.
▶ Only operate devices or components if they are undamaged on the outside.
▶ Check the power cable before use.
▶ Connect only to sockets with a protective contact that meet the respective
national regulations.
▶ Check the proper working order and proper condition of product and ac-
cessories before each use.
▶ Have parts with sites of breakage or surface changes checked by the Ser-
vice.
▶ Safety checks may only be performed by trained service personnel.
▶ Perform a test run with the handpiece before each use.
▶ If any of the following defects on the product or accessories occurs, stop
working and have the service personnel carry out repair work:
▪ Malfunctions
▪ Damage
▪ Irregular running noise
▪ Excessive vibration
10 / 55