12 Information about electromagnetic compatibility | 12.3 Guidelines and manufacturer's statement - Electromag-
Interference im-
IEC 60601 test
munity tests
levels
Electrostatic discharge
± 8 kV contact dis-
(ESD) according to
charge
IEC 61000-4-2
± 15 kV atmospheric
discharge
Fast transient elec-
± 2 kV for power lines
trical interference /
± 2 kV for input and
bursts according to
output lines
IEC 61000-4-4
Surges according to
± 1 kV push-pull
IEC 61000-4-5
voltage
± 2 kV common mode
voltage
Voltage interruptions,
0 % / 0.5 cycles
short-term interrup-
at 0º to 315º in 45º
tions, and fluctuations
increments
of the supply voltage
0 % / 1 cycle
according to IEC
70 % / 25 cycles
61000-4-11
0 % / 250 cycles
Magnetic field at a
30 A/m
supply frequency
(50/60 Hz) according
to IEC 61000-4-8
12.3 Guidelines and manufacturer's statement -
Electromagnetic immunity
The elements e-motion is intended for use in an environment of the kind spe-
cified below. The customer or user of the elements e-motion should make sure
that the device is used in an environment of the specified type.
Instructions for use elements e-motion Dental Motor
Compliance level
Electromagnetic environment
- Guidelines
± 8 kV contact dis-
Floors should be made of wood
charge
or concrete or be fitted with
± 15 kV atmospheric
ceramic tiles. If the floor is fitted
discharge
with synthetic material, the relat-
ive humidity must be at least 30
%.
± 2 kV for power lines
The quality of the supply voltage
and input and output
should correspond to that of a
lines
typical business or hospital envir-
onment.
± 1 kV push-pull
The quality of the supply voltage
voltage
should correspond to that of a
± 2 kV common mode
typical business or hospital envir-
voltage
onment.
0 % / 0.5 cycles
The quality of the supply voltage
at 0º to 315º in 45º
should correspond to that of a
increments
typical business or hospital envir-
0 % / 1 cycle
onment.
70 % / 25 cycles
0 % / 250 cycles
30 A/m
Magnetic fields at the mains fre-
quency should correspond to the
typical values in a business and
hospital environment.
53 / 55