Instructions for use elements e-motion Dental Motor
8 Reprocessing steps in accordance with DIN EN ISO 17664 | 8.5 Storage
▪ elements 8:1 handpiece
▪ INTRA LUX Motor KL 703 LED
Note
The reprocessing steps for the elements 8:1 handpiece or the INTRA LUX
motor KL 703 LED are described in the corresponding Instructions for use.

8.5 Storage

Prepared products must be stored appropriately in a dry, dark, cool room such
that they are protected from germs (as far as possible) and dust.

8.6 Service, inspection and testing after preparation

Note
It is essential to comply with the hygiene requirements (sterility) during the
test after reprocessing. If sites of fracture and clear changes of the surface
are visible, the parts need to be checked by the Service.
Check for cleanliness, intactness, servicing, and repair as described in the fol-
lowing:
▶ Check the adjustable functions of the unit and the motor function.
▶ Check the control commands on the foot control.