Instructions for use elements e-motion Dental Motor
10 Troubleshooting
Malfunction
Event E5
Event E8
Event E12
Event E14
All other events
Cause
Automatic light and
motor shut-off during
continuous operation of
the motor.
Saved data or settings
have been reset to
starting value.
No connection to mo-
tor.
Motor overload.
Internal system error.
50 / 55
Remedy
▶ Comply with defined operating mode.
▶ Confirm message and check or correct the pro-
gram settings.
▶ If error persists, notify the service engineer.
▶ Check/restore correct connection.
▶ Replace motor cord.
▶ If error persists, notify the service engineer.
▶ Relieve the motor, stop and start on foot con-
trol.
▶ Turn the unit off and on.
▶ If error persists, notify the service engineer.