Instructions for use elements e-motion Dental Motor
8 Reprocessing steps in accordance with DIN EN ISO 17664 | 8.1 Cleaning
8 Reprocessing steps in accordance with DIN EN ISO
17664
Note
The reprocessing steps for the motor and the straight and contra-angle hand-
pieces are described in the corresponding instructions for use.
Improper disinfection.
Property damage to the product.
▶ Do not immerse product in liquids.
▶ Use disinfectant in accordance with the instructions of the manufacturer.
▶ Do not use spray disinfection.
▶ Only disinfect by wiping.
▶ Never use chlorine-containing disinfectants.
Note
The reprocessing instructions have been validated by the manufacturer. Any
departure from the instructions provided must be checked by the repro-
cessing entity for efficacy and possible detrimental consequences.
8.1 Cleaning
Note
Do not use solvents or aggressive chemicals.
8.1.1 Preparations at the site of use
▶ Unplug the unit from the mains.
▶ Decontaminate as soon as possible after application.
▶ Extreme contamination must be removed right after it is generated.
8.1.2 Manual external cleaning
Note
Do not use scouring cleansers.
▶ Make sure that the device is disconnected from the mains.
▶ Dampen a soft cloth with tap water or a mild cleaning solution (weak soapy
water).
▶ Wipe all external surfaces of the elements e-motion housing and the ex-
ternal surfaces of the motor cord with the lightly dampened cloth.
8.1.3 Manual internal cleaning
There is no specific cleaning of the inside of the unit.
8.1.4 Automated external and internal cleaning
The following parts of the unit are released for machine-based cleaning:
▪ elements 8:1 handpiece
NOTICE