Instructions for use elements e-motion Dental Motor
6 Operation | 6.4 Operating the elements e-motion
This file has a torque of, e.g., 4.00 Ncm assigned to it.
▶ If the speed is active, activate the torque by touching the torque value.
▶ Tap the "Lock" control element to unlock settings.
▶ If the settings are unlocked, pull the slider to the right or left to change the
torque.
The torque is limited to the set value.
Note
A notice signal is issued once a level of 90% of the set torque is reached.
If the set torque of the file differs from the recommended value of this file, a
crossed-out factory symbol is shown.
6.4.3 Selecting torque mode
Note
The torque mode is deactivated in the TF Adaptive Motion.
There are three different torque modes:
▪ Autorev/Forward
▪ Auto-reverse
▪ Torque Control
▶ Tap the "Lock" control element to unlock settings.
▶ If the settings are unlocked, press the "Torque mode" control element to
display a selection list.
▶ Select the torque mode from the selection list.
ð
The torque mode is shown on the display.
ð
The displayed torque mode is saved.
32 / 55