OPERATION
3.2.

Commissioning the instrument

Before the instrument can be used for monitoring possible toxic gases in
the atmosphere, it must be commissioned.
(1) Press and hold [TEST] key for 3 s, until "
(2) Release [TEST] key and press again.
After about 1 s, all display indicators are activated;
Acoustic alarm sounds, alarm LEDs light up and vibration alarm is
triggered;
The software version is displayed for approx. 3 s;
The gas type is displayed for approx. 3 s more (O
LO
approx. 3 s.
During the display of
corresponding setpoints can still be changed ( Section 3.1).
(3) Release [TEST] key and wait for 3 s.
Instrument begins an activation cycle of 99 s duration (countdown).
After activation, the remaining operation time is displayed in months.
The ready indicator lights up every 60 s and the alarm LEDs light up
briefly.
After successful activation, the instrument remains active without
interruption until the battery is flat.