7.1.1 Measuring mode
Display example
Description
In normal operation, the display
20.9
shows the measured gas concentra-
tion, the selected gas and the unit of
measurement.
O 2
Vol.%
Exceeding the alarm thresholds.
19.0
Example shows pre alarm A1.
O 2
A1
7.1.2 Special states
If the instrument is in a special state, a proper measurement
or alarming can no longer be guaranteed.
This is the case:
– When the gas concentration exceeds the measuring
range.
– When a fault occurs.
– When alarms are deactivated.
The measuring function (display output and analog/digital
interface) remains active.
– During calibrations.
– During bump tests.
– During warm-up phase.
– During maintenance.
For the current of the analog interface corresponding to the
analog signal, see 16.1.
Display example
Description
Value is over measuring range
The gas concentration is beyond the
measuring range of the sensor.
Analog interface: Measuring range
O 2
A2
exceeded
Relays: A2 relay switches
Value is under measuring range
The gas concentration is beyond the
measuring range of the sensor.
Analog interface: Drift below zero
O 2
Vol.%
Relays: Fault relay switches
Fault indication
The
side of the display.
Analog interface: Fault current
Relays: Fault relay switches
Maintenance indication
20.9
The
side of the display.
This is displayed when alarms are
O 2
Vol.%
deactivated, during calibrations, bump
test and maintenance work.
Analog interface: Maintenance cur-
rent
Relays: No change
Instructions for use
|
Dräger PointGard 2000 Series
symbol is displayed on the right
symbol is displayed on the right
Display example
Description
Warning indication
20.9
The
side of the display.
Analog interface: Warning current
O 2
Vol.%
Relays: No change
1) Only when warning current is enabled. Factory default setting: dis-
abled.
Warm-up phase 1
The
symbol is displayed on the right side of the display.
The remaining time is indicated on the left side of the display.
Analog interface: Maintenance current
Relays: Fault relay switches
For warm-up phase 1, the fault relay behavior is configurable
(see "Setting fault relay warm-up 1", page 23)
Warm-up phase 2
The
symbol is displayed on the right side of the display.
The measured value is indicated on the left side of the display.
Analog interface: Measured value
Relays: Behavior as in measuring mode (Fault relay does not
indicate warm-up phase 2.)
7.1.3 Ending special states
Value under/over measuring range
The display indicates that the gas concentration is beyond the
measuring range of the sensor. The indication disappears as
soon as the gas concentration is within the measuring range.
NOTICE
Sensor malfunction
Gas concentrations beyond the measuring range of the
sensor might cause sensor malfunctions.
► Check the calibration.
PointGard 2200 CAT-specific:
An over range has to be acknowledged with [OK] after
verifying that the gas concentration is below 100 %LEL (e.g.
using a portable instrument).
Latching alarms is the default setting of the instrument.
For instruments without relays, the over range can be set to
non-latching. In this case, the over range indication
disappears. The control unit must support this function! For
further information see: "Setting DD sensor latching",
page 30.
Fault / warning indication
Faults and warnings are non-latching. If the fault or warning
condition clears, the message disappears.
To solve warning and fault conditions, display the error code
or warning message (see "Displaying Information", page 17)
and begin troubleshooting (see "", page ).
Operation
symbol is displayed on the right
|
en
1)
15