17
Technical Information
Connection of the High Vacuum Flange
Figure 54
Use the required number of claw clamps:
Tighten the claw clamps with a torque of 22 Nm.

Installation of ConFlat Flange

For ConFlat flange connections, fix the two flanges with the screws
as shown in the following figures (a protective screen can optionally
be used).
292/304
4 for ISO 100 flange
4 for ISO 160 flange
TwisTorr 304 FS User Manual / 87-901-027-01(B)