Descargar Imprimir esta página

Interconnections; Earth (Ground) Connection - Agilent Technologies TwisTorr 304 FS Serie Manual De Istrucciones

Ocultar thumbs Ver también para TwisTorr 304 FS Serie:

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 57
17
Technical Information

Interconnections

Interconnections

The following figure shows the TwisTorr 304 FS interconnections.
Figure 32

Earth (Ground) Connection

We recommend that you fit a separate earth (ground) conductor to
the TwisTorr 304 FS. Use a separate insulated green/yellow
conductor, and use a M5 x 10 screw and shake proof washer (fitted to
the earth hole on the pump) to secure the earth conductor to the
pump. The screw must be fixed with a 2 Nm torque.
The impedance between the pump-body and the earth connection
point must be < 0.1 Ω.
270/304
TwisTorr 304 FS User Manual / 87-901-027-01(B)

Publicidad

loading