Descargar Imprimir esta página

Technical Specification - Agilent Technologies TwisTorr 304 FS Serie Manual De Istrucciones

Ocultar thumbs Ver también para TwisTorr 304 FS Serie:

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 57
17
Technical Information

Technical Specification

Technical Specification

Max gas throughput:
water cooling (water temperature=25°C; backing pump 5 m
o
air cooling (ambient temperature 25°C; backing pump 5 m
o
Tab. 3 Technical Specification
Pumping speed
N
2
He
H
2
Ar
Max gas throughput at full rotational speed
(with recommended forepump)
N
2
Ar
Compression ratio
N
2
He
H
2
Ar
Base pressure with recommended forepump
Inlet flange
Foreline flange
Rotational speed
Start-up time
Recommended forepump
248/304
ISO 100 / CF 6"
250 l/s
255 l/s
220 l/s
250 l/s
Air cooling
170 sccm
110 sccm
>1  10
11
>1  10
8
1.5  10
6
> 1  10
11
 1 x 10
mbar
-10
( 1 x 10
Torr)
-10
(According to standard DIN 28 428, the base pressure is
that measured in a leak-free test dome, 48 hours after the
completion of test dome bake-out, with a Turbopump fitted
with a ConFlat flange and using the recommended pre-
vacuum pump)
CFF 8" O.D.
ISO 160
CFF 6" O.D.
ISO 100
KF16 NW (KF25 – optional)
60000 rpm (1010 Hz driving frequency)
 3 minutes
mechanical: Agilent DS 102
mechanical: Agilent DS 302
dry pump: Agilent SH 110
TwisTorr 304 FS User Manual / 87-901-027-01(B)
/h)
3
/h)
3
ISO 160 / CF 8"
250 l/s
255 l/s
220 l/s
250 l/s
Water cooling
170 sccm
110 sccm

Publicidad

loading