Technical Information
Window-Meanings
N.
107
108
109
110
111
117
120
122
125
126
308/318
Read/
Data
Description
Write
Type
R/W
L
Active Stop
(write only in stop)
R/W
N
Baud rate
W
L
Pump life/ cycle time/ cycle
number reset
R/W
L
Interlock type
(default = 1)
R/W
L
Analog output type: output
voltage signal proportional
to frequency or power
R/W
N
Low Speed frequency (Hz)
R/W
N
Rotational frequency setting
(Hz)
R/W
L
Set vent valve on/off
R/W
L
Set the vent valve operation
R/W
N
Vent valve opening delay
(expressed in 0.2 sec)
Admitted Values
0 = NO
1 = YES
(default = 0)
600 = 0
1200 = 1
2400 = 2
4800 = 3
9600 = 4
(default = 4)
To reset write '1'
Impulse = 0
Continuous = 1
0 = frequency
1 = power
(default = 1)
2 = Pump temperature
3 = Pressure
1100 to "Maximum rotational
frequency"
(win 120)
(default = 1100)
1100 to 1167 (default = 1167)
0 = Closed
1 = Open
(default = 1)
Automatic = 0 (see note 1.)
On command = 1 (see note 2.)
0 to 65535
(corresponding to 0 to 13107
sec) (default = 15)
TPS-mini User Manual / 87-901-003-01 (E)