Agilent Technologies TPS-mini Serie Manual De Instrucciones página 311

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 61
N.
500
503
504
1. Automatic means that when the controller stops, the vent valve is opened
NOTE
with a delay defined by window n. 126; when the controller starts, the vent
valve is immediately closed.
2. On command means that the vent valve is opened or closed by means of
window n. 122.
3. These commands remove the pump motor supply for a short time. If they are
used continuously, they can cause the pump stopping.
Figure 35
TPS-mini User Manual / 87-901-003-01 (E)
Read/
Data
Description
Write
Type
Reserved to Agilent service
R/W
N
RS 485 address
R/W
L
Serial type select
7
6
TOO HIGH LOAD
SHORT CIRCUIT
OVERVOLTAGE
Window N. 206 Bit Description
5
4
3
2
1
0
NO CONNECTION
PUMP OVERTEMP.
CONTROLL. OVERTEMP.
Technical Information
Window-Meanings
Admitted Values
0 to 31
(default = 0)
0 = RS 232
1 = RS 485
(default = 0)
311/318
17

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

969918096991819699182

Tabla de contenido