Agilent Technologies TwisTorr 84 FS AG Serie Manual De Instrucciones página 163

Ocultar thumbs Ver también para TwisTorr 84 FS AG Serie:
Tabla de contenido

Publicidad

PROGRAMMABLE ANALOG SIGNAL: this output signal is a voltage
(from 0 to 10 Vdc) proportional to a reference quantity (frequency or
power or Temperature or Pressure) set by the user through serial
communication using Win 111 (see below table). The default setting
is the frequency (see the following example diagram).
Figure 16
Type
Frequency
Power
Temperature
Pressure
TwisTorr 84 FS AG Navigator Controller User Manual / 87-901-038-01 (A)
10
9
8
7
6
5
4
3
2
1
0
0
150
300
450
Analog Output Voltage proportional to frequency (WIN 111=0)
Range
0 ÷ 1350Hz
0 ÷ 100W
0 ÷ 100 °C
1*10
[Torr] ÷ 1*10
-9
-1
Technical Information
600
750
900 1050 1200 1350
Frequency [Hz]
Output Analog Voltage
0 ÷ 10Vdc
0 ÷ 10Vdc
0 ÷ 10Vdc
[Torr]
0 ÷ 10Vdc
17
J1 – IN-OUT
WIN 111
0
1
2
3
163/192

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

X3509-64000X3509-64001

Tabla de contenido