Pump Connector - Agilent Technologies TwisTorr 84 FS AG Serie Manual De Instrucciones

Ocultar thumbs Ver también para TwisTorr 84 FS AG Serie:
Tabla de contenido

Publicidad

Technical Information
Controller Connection

Pump Connector

The controller is equipped with a 30-cm long pump cable. The
following figure shows the controller output connector configuration
with pins:
Figure 12
If the temperature sensor is disconnected, the pump will not start. To
prevent damage to the pump when the temperature exceeds 60° C,
the sensor automatically cuts off the power supply.
158/192
A-F = pump temperature sensor
B-C-D = 60 Vac 3-phase output to pump motor stator
E = EARTH PROTECTION
Controller-to-Pump Connector
TwisTorr 84 FS AG Navigator Controller User Manual / 87-901-038-01 (A)
Earth

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

X3509-64000X3509-64001

Tabla de contenido