Compex performance Manual Del Usuario página 310

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

RECOMMENDATIONS AND DECLARATION BY THE MANUFACTURER -
Performance must ensure it is used in this recommended environment.
Electrostatic discharge (DES)
Rapid electrical transients in
Voltage trough, short power
cuts and voltage variations on
power supply input lines
Magnetic field at mains
frequency (50/60 Hz)
NOTE: VT is the AC supply voltage before application of the test
level.
308
The Compex Performance is designed for use in the electromagnetic
environment stipulated below. The buyer or user of the Compex
Immunity test
IEC 61000-4-2
bursts of noise
IEC 61000-4-4
Surge currents
IEC 61000-4-5
IEC 61000-4-11
IEC 61000-4-8
Test level IEC 60601
± 6 kV at the contact
± 8 kV in air
± 2 kV for power supply
lines
± 1 kV for input/output lines
± 1 kV differential mode
± 2 kV common mode
< 5% VT (trough > 95% of
VT) for 0.5 cycle
< 40% VT (trough > 60% of
VT) for 5 cycles
< 70% VT (trough > 30% of
VT) for 25 cycles
< 5% VT (trough > 95% of
VT) for 5 seconds
3 A/m
Observance level
± 6 kV at the contact
± 8 kV in air
Not applicable
System battery-powered
Not applicable
System battery-powered
Not applicable
System battery-powered

Publicidad

Tabla de contenido
loading

Productos relacionados para Compex performance

Tabla de contenido