Technical Documentation (Con't)
XXXIII
IEC 60601-1-2, Table 202

Electromagnetic Immunity

Invia Liberty is intended for use in the electromagnetic environment specified below. The cu-
stomer or the user of Invia Liberty should assure that they are used in such environment.
Immunity Tests
Electrostatic
Discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
short interrup-
tions and voltage
variations on
power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
199
IEC 60601
Compliance
test level
level
± 6 kV contact
± 6 kV contact
± 8 kV air
± 8 kV air
± 2 kV
± 2 kV
for power
for power
supply lines
supply lines
± 1 kV
± 1 kV
for input /
for input /
output lines
output lines
± 1 kV
± 1 kV
differential mode
differential mode
± 2 kV
± 2 kV
common mode
common mode
<5 % UT
<5 % UT
(>95 % dip in UT)
(>95 % dip in UT)
for 0.5 cycle
for 0.5 cycle
40 % UT
40 % UT
(60 % dip in UT)
(60 % dip in UT)
for 5 cycles
for 5 cycles
70 % UT
70 % UT
(30 % dip in UT)
(30 % dip in UT)
for 25 cycles
for 25 cycles
<5 % UT
<5 % UT
(>95 % dip in UT)
(>95 % dip in UT)
for 5 sec
for 5 sec
3 A/m
3 A/m
Electromagnetic
environment - guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidi-
ty should be at least 30 %.
Mains power quality should be that of
a typical commercial or hospital envi-
ronment.
Mains power quality should be that of
a typical commercial or hospital envi-
ronment.
Mains power quality should be that of
a typical commercial or hospital en-
vironment. If the user of Invia Liberty
requires continued operation during
power mains interruptions, it is re-
commended that Invia Liberty be po-
wered from an uninterruptible power
supply or a battery.
Power frequency magnetic fields
should be at levels of a typical com-
mercial or hospital environment.