Option
CLEAN INSTRUMENT
CONFIGURE NH2CL AS N OR
CL2
SIGNAL AVERAGE
EDIT ANALYZER NAME
EDIT CHANNEL NAME
CONFIGURE SEQUENCER
(optional)
SET DATE & TIME
DISPLAY SETUP
DISABLE REMINDERS
MANAGE DEVICES
INSTRUMENT INFORMATION
CALCULATION
SETUP OUTPUTS
SETUP NETWORK (optional)
SECURITY SETUP
RESET DEFAULTS
DECOMMISSIONING
Manage devices
Install or remove input modules.
1. Push menu and select MANAGE DEVICES.
2. Select an option.
Option
SCAN FOR DEVICES The system shows the connected devices.
DELETE DEVICE
Description
Sets the cleaning interval for the analyzer. Immediately starts a manual
cleaning or the scheduling of an automatic cleaning. Set the cleaning mode
to occur daily at a selected time, on a selected day of the week at a
selected time or after a selected number of hours (range: 2 to 999 hours).
Changes the NH
Cl
.
2
Selects the number of measurements used to calculate a moving average
measurement. This reduces variability in measurements (default = 1, no
averaging).
Changes the name that is shown on the top of the measurement screen
(maximum of 16 characters).
Changes the name of the sample source that is shown on the
measurement screen (maximum of 10 characters).
Starts or stops measurements for individual sample sources. Sets the
measurement order of the sample sources when there is more than one
sample source.
Sets the analyzer time and date.
Changes the language. Adjusts the order measurements are shown.
Adjusts the display contrast settings.
Stops scheduled maintenance alerts for individual components. Options:
tubing, stir bar, colorimeter cells, air filter, air relief valve, pinch valve, air
compressor, reagent valves, sample pump, fan filter, air check valve,
colorimeter check valve, internal filter and distribution valve.
Installs or removes input modules. Refer to
for more information.
Shows the analyzer information. Refer to
on page 17.
Set up variables, parameters, units and formulas for the analyzer. Refer to
Set up the calculation
Selects and configures the 4–20 mA setup, the relay setup and the error
hold mode. Refer to
information.
Shows only if a network card is installed. The network card support is for
Modbus and RS485.
Enables or disables the pass code (default = HACH55).
Sets the configuration to the factory defaults.
Starts the cycle to prepare the analyzer for storage. Refer to the
maintenance and troubleshooting documentation for more information.
Description
Note: If no device is connected, the system returns to the main measurement
screen.
Removes the device when the device is no longer connected.
Cl value shown on the measurement screen as N or as
2
on page 12.
Configure the outputs
Manage devices
on page 11
View instrument information
on page 12 for more
English 11