Configure The Sequencer (Optional); Calibrate The Analyzer; Operation; Set Up The Reagents And Standards - Hach Ammonia Monochloramine 5500sc Operaciones

Ocultar thumbs Ver también para Ammonia Monochloramine 5500sc:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Configure the sequencer (optional)

Option only available for the multi-channel version of the analyzer.
1. Select SETUP SYSTEM>CONFIGURE SEQUENCER.
Option
ACTIVATE
CHANNELS
SEQUENCE
CHANNELS

Calibrate the analyzer

The manufacturer recommends calibration of the analyzer after 1 day of operation to allow all system components
to stabilize.
Refer to
Calibration
on page 19 to start a calibration.

Operation

Potential fire and explosion hazard. This equipment is intended for only aqueous type samples. Use with
flammable samples can result in a fire or explosion.
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective
equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets
(MSDS/SDS) for safety protocols.
When the lower door is opened, the current measurement or calibration is stopped. The
measurement or calibration results will be discarded. When the lower door is closed, the stopped
operation starts again.
Note: Configure the analog outputs and relay behavior for this interruption time. Refer to
on page 16 to select HOLD or TRANSFER. Then, refer to
setup
on page 14 for the SET TRANSFER settings.

Set up the reagents and standards

Make sure to set the sample flow rate and install the reagent bottles before this task is started.
1. Push menu and then go to REAGENTS/STANDARDS.
2. Select one of the options. Use the SET options when the volume of existing reagents/standards
needs to be adjusted. Use the RESET options when reagents/standards are replaced or renewed
(bottles are replaced).
Option
SET REAGENT LEVEL
Description
Starts or stops measurements for individual sample sources. Use the UP and DOWN
navigation keys to scroll through the channels. Deselect a channel with the LEFT
navigation key. Push enter to confirm.
Note: Inactive channels are shown with a "~" character before the channel name on all
displays.
Sets the measurement order of the sample sources. Use the UP and DOWN navigation
keys to scroll through the sequence. For each sequence number use the LEFT and
RIGHT navigation keys to select a channel. Push enter to confirm.
N O T I C E
W A R N I N G
Description
Sets the volume of a specific reagent in the reagent bottle to an estimated
value. Range: 1–100%.
C A U T I O N
4–20 mA module setup
Set the error hold mode
on page 12 and/or
Relay
English 9

Publicidad

Tabla de contenido
loading

Tabla de contenido