Calibration - Hach Ammonia Monochloramine 5500sc Operaciones

Ocultar thumbs Ver también para Ammonia Monochloramine 5500sc:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

4. When the upgrade is complete, the display shows TRANSFER COMPLETE. Remove the SD
card.
5. Restart the instrument for the upgrade to take effect.

Calibration

The manufacturer recommends calibration of the analyzer after 1 day of operation to allow all system components
to stabilize.
The automatic calibration uses the installed known standards to calibrate the analyzer. Calibration
can be done manually or scheduled to occur automatically. Use the calibration menu to view
calibration data, start or cancel the auto calibration, override the auto cal settings, calibrate manually
or to reset to the default calibration.
1. Push cal to display the calibration menu.
Option
START MANUAL CAL
SET AUTO CALIBRATION Refer to
CALIBRATION DATA
OVERRIDE CALIBRATION To calibrate the slope value, refer to the concentration value from the user
OUTPUT CALIBRATION
RESET DEFAULT CAL
Schedule automatic calibrations
1. Push cal, then select SET AUTO CALIBRATION.
2. ENABLE AUTO CAL>YES.
3. Select STD SOLUTION and enter the standard value in ppm.
4. Select a schedule option for calibration.
Option
Description
TIME BASE
Sets the interval between calibrations. Options: DAYS or HOURS.
WEEK DAY
Selects the day or days of the week for calibration when TIME BASE is set to DAY.
TIME
Sets the time of day for calibration when the TIME BASE is set to DAY.
SET INTERVAL Sets the interval between automatic calibrations in hours when the TIME BASE is set to
HOURS (range: 2 to 999 hours).
Start a calibration manually
1. Push cal, then select START MANUAL CAL.
2. The measurement status is displayed. Select YES to interrupt the current measurement cycle and
start the calibration immediately. Select NO to wait until the current measurement cycle has
completed before the calibration starts.
3. Complete the instructions on the display.
N O T I C E
Description
Refer to
Start a calibration manually
Schedule automatic calibrations
Displays data from the last calibration and the due date and time of the next
scheduled calibration.
input. When override calibration is selected, the automatic calibration is
disabled.
Select a 4-20 mA output and enter the output values to be sent.
Reset calibration data to default values and disables automatic calibration. On
completion, perform a new calibration.
on page 19.
on page 19.
English 19

Publicidad

Tabla de contenido
loading

Tabla de contenido