Hach Ammonia Monochloramine 5500sc Operaciones página 13

Ocultar thumbs Ver también para Ammonia Monochloramine 5500sc:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

3. Select SET FUNCTION>LINEAR CONTROL and then select the applicable options in the
ACTIVATION menu.
Option
SET LOW VALUE
SET HIGH VALUE
4. Select SET FUNCTION>PID CONTROL and then select the applicable options in the
ACTIVATION menu.
Option
Description
SET MODE
AUTO—the signal is automatically controlled by the algorithm when the analyzer uses
proportional, integral and derivative inputs.
MANUAL—the signal is controlled by the user. To change the signal manually, change the
% value in MANUAL OUTPUT.
PHASE
Selects the signal result when process changes occur.
DIRECT—signal increases as the process increases.
REVERSE—signal increases as the process decreases.
SET SETPOINT Sets the value for a control point in the process.
PROP BAND
Sets the value for the difference between the measured signal and the necessary setpoint.
INTEGRAL
Sets the period of time from the reagent injection point to the contact with the measuring
device.
DERIVATIVE
Sets a value that adjusts for vacillation of the process. The majority of applications can be
controlled without the use of the derivative setting.
TRANSIT TIME
Sets the value to stop the PID control for a selected period of time when the sample moves
from the control pump to the measurement sensor.
5. Select SET FUNCTION>LOGARITHMIC and then select the applicable options in the
ACTIVATION menu.
Option
SET 50% VALUE
SET HIGH VALUE
6. Select SET FUNCTION>BILINEAR and then select the applicable options in the ACTIVATION
menu.
Option
SET LOW VALUE
SET HIGH VALUE
SET KNEE POINT VALUE
SET KNEE POINT CURRENT Sets the value of the current at the knee point value.
Description
Sets the low endpoint of the process variable range.
Sets the high endpoint (upper value) of the process variable range.
Description
Sets the value corresponding to 50% of the process variable range.
Sets the high endpoint (upper value) of the process variable range.
Description
Sets the low endpoint of the process variable range.
Sets the high endpoint (upper value) of the process variable range.
Sets the value at which the process variable range divides into another linear
segment.
English 13

Publicidad

Tabla de contenido
loading

Tabla de contenido