Set up the calculation
Set up variables, parameters, units and formulas for the analyzer.
1. Push menu and select CALCULATION.
2. Select an option.
Option
SET VARIABLE X
SET PARAMETER X Selects the parameter referring to the variable X.
SET VARIABLE Y
SET PARAMETER Y Selects the parameter referring to the variable Y.
SET FORMULA
DISPLAY FORMAT
SET UNITS
SET PARAMETER

Configure the outputs

4–20 mA module setup
1. Push menu and select SETUP SYSTEM>SETUP OUTPUTS>4–20mA SETUP.
2. Select the OUTPUT.
3. Select an option.
Option
ACTIVATION
SELECT SOURCE Select the output. Options: None if the output is not configured, the analyzer name or
SET PARAMETER Select the measurement channel from the list.
SET FUNCTION
SET TRANSFER
SET FILTER
SCALE 0mA/4mA
4–20 mA activation options
1. Push menu and select SETUP SYSTEM>SETUP OUTPUTS>4–20mA SETUP.
2. Select the applicable OUTPUT.
12 English
Description
Selects the sensor referring to the variable X.
Selects the sensor referring to the variable Y.
Selects the calculation formula to complete. Options: None, X-Y, X+Y, X/Y, [X/Y]%, [X
+Y]/2, X*Y, [X-Y]%/X
Selects the number of decimal places shown in a calculation result. Options: Auto,
XXXXX, XXXX.X, XXX.XX, XX.XXX, X.XXXX
Enters the unit name (maximum of 5 characters).
Enters the measurement name (maximum of 5 characters).
Description
The menu list items change with the selected function. Refer to
options
on page 12 for more information.
calculation if a calculation formula has been configured. Refer to
on page 12.
Select a function. Further options will vary depending on which function is chosen.
LINEAR CONTROL—Signal is linearly dependent on the process value. PID CONTROL
—Signal works as a PID (Proportional, Integral, Derivative) controller. LOGARITHMIC—
Signal is represented logarithmically within the process variable range. BILINEAR—
Signal is represented as two linear segments within the process variable range.
If TRANSFER is or will be selected as the ERROR HOLD MODE, select SET
TRANSFER and enter the transfer value. Range: 3.0 to 23.0 mA (default = 4.000). Refer
to
Set the error hold mode
Enter the filter value. This is a time-average filter value of 0 to 120 seconds (default = 0).
Select the scale (0–20 mA or 4–20 mA).
on page 16.
4–20 mA activation
Set up the calculation