9. Technische beschrijving

Item
Stroombron
DC12V, 1,5 A met adapter
Stroomverbruik
12V DC, 850mA
Afmetingen apparaat (L x B x H)
9.8" x 4.7" x 2.2" of 25 x 12.2 x 5.7cm
Gewicht apparaat
1.0 kg of 2.2lb (batterij en adapter inbegrepen)
Werking: 5°C tot 40°C (41°F tot 104°F)
opmerking)
Temperatuur
Bewaring: -10°C tot 50°C (14°F tot 122°F)
Omgeving
Verzending: -10°C tot 70°C (14°F tot 158°F)
Werking : 30% tot 75% niet-condenserend
Vochtigheid
Opberging: 10% tot 75% niet-condenserend
Verzending: 10% tot 90% niet-condenserend
Type BF, IPX0
Toegepast deel: luchtkussen
Classificatie
Niet geschikt voor gebruik in aanwezigheid van ontvlambare
anesthetische mengsels
(Geen AP- of or APG-bescherming)
Drukbereik
Keuze uit vijf instellingen
Cyclustijd
10, 15, 20 minuten
6V Ni-MH batterij (typisch 2000mAh )
Battery pack
(aanbevolen type NEXcell , LSD-AAx5 )
Kussen
Afmeting (L x B x H)
17" x 17" x 4" of 43 X 43 X 10 cm
Gewicht
1,1 kg of 2,4lb
Opmerking:
1. Het batterijvermogen kan gereduceerd worden als het apparaat in een koude omgeving gebruikt
wordt.
2. Raadpleeg de verdeler of een EG-vertegenwoordiger voor andere technische documenten .
3. Gelieve de nationale vereisten voor de reglementaire verwijdering van de eenheid te volgen .
4. De fabrikant behoudt zich het recht voor om de technische kenmerken van het product zonder
voorafgaand bericht te wijzigen.
Technisch kenmerk
(* Zie onderstaande
Technisch kenmerk
80
English

8. Troubleshooting

Problem
‧ Check if the plug is connected to mains.
‧ Check if the power cord is well connected to the pump
Power is not ON
‧ To operate from battery, check if the DC plug is
disconnected from the pump unit
‧ Check if the connection between air tube connector to
pump unit is tightly secured.
‧ Check if all tubing connections along cushion are
Alarm is on
(audible & visual)
secured.
‧ Check if there is any leakage from air cells.
‧ Check if the battery is low
‧ The pressure is too firm for too long, release some air by
Cushion becomes too firm and
disconnecting the air tube connector and change to your
pressure can't be lowered
preferred setting.
‧ This is normal since it is at alternating mode. Air outlets
No air is produced from some air
outlet of the air tube connector
take turns to produce air during their preset cycle time.
‧ Check each air cell. Please select the operation mode as
static mode so that all the air cells will be inflated. Then
ensure each single cell is not broken.
‧ Examine for air leakage from tubes. Ensure no leakage
Low pressure, unable to resolve.
occurs.
‧ If any leakage occurs, please contact your vendor.
‧ Plug in the pump unit and check the airflow from the
hose connection port. The airflow should alternate
between ports every half-cycle.
‧ After the battery is fully charged (about 12 hours), the
pump immediately has 8 hours of continuous operation
time. If it is unable to operate over 8 hours, the battery
Battery unable to obtain full
pack might need to be replaced.
‧ If the low battery indicator still lights up after it has been
charge.
properly recharged for approximately 12 hours, the
battery pack might need to be replaced. Replacement
required after about 6 months.
If your questions are not answered by above information, please contact your local agent directly.
They might require a technician to take care of the problem.
Solution