Technical Data; English - Milwaukee 2231-20 Manual Original

Ocultar thumbs Ver también para 2231-20:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 9

TECHNICAL DATA

Range
Resolution
0.00 mA ‑ 3.99 mA
0.01 mA
4.00 mA ‑ 21.49 mA
0.01 mA
21.5 mA ‑ 99.9 mA
0.1 mA
Accuracy is specified for 1 year after calibration, at operating temperatures of 18°C to 28°C, with relative humidity at 0 % to 85 %.
Jaw Opening (maximum conductor size)
Temperature Operating
Temperature Storage
Temperature Coefficient
Operating Altitude
Drop Test
Battery
Battery Life
Weight according EPTA‑Procedure 01/2003
Safety Compliances ....EN61010‑1, UL 61010‑1, IEC 61010‑2‑32 (Clamp Assemblies), IEC/EN 61010‑1 2nd Edition for measurement
Category II, 60 V, Pollution Degree 2, EMC EN61326‑1
Certifications ...cULus, CE
WARNING! Read all safety warnings and all
instructions, including those given in the accompanying
brochure. Failure to follow the warnings and instructions may
result in electric shock, fire and/or serious injury.
Save all warnings and instructions for future reference.
SAFETY INSTRUCTIONS Measuring Instrument
To avoid electrical shock: Use only for tests on 60V Cat II DC
circuits. Do not use to test AC current.
Do not make tests on non‑insulated conductors.
Transformer jaw tips are designed not to short the circuit under
test. If equipment under test has exposed conductive parts,
however, extra precaution should be taken to minimize the
possibility of shorting.
Check meter functionality on a known circuit first. Never
assume meter is working. Assume circuits are live until they
can be proven de‑energized.
Never open the battery cover during a measurement.
The instrument is to be used only in its intended applications or
conditions. Otherwise, safety functions equipped with the
instrument doesn't work, and instrument damage or serious
personal injury may be caused.
To reduce the risk of injury from shock and arc blasts, always
wear personal protective equipment where live conductors are
exposed.
Verify proper operation on a known source before use or taking
action as a result of the indication of the instrument.
Do not use with the battery cover removed.
Do not attempt to make measurement in the presence of
flammable gasses. Otherwise, the use of the instrument may
cause sparking, which can lead to an explosion.
Never attempt to use the instrument if its surface or your hand
is wet. Do not exceed the maximum allowable input of any
measuring range.
Do not ground yourself while measuring. Avoid body contact
with earthed or grounded sur‑ faces such as pipes, radiators,
ranges and refrigerators.
Never attempt to make measurement if any abnormal
conditions, such as broken case and exposed metal parts are
found on the instrument.
Do not install substitute parts or make any modification to the
instrument. For repair or re‑ calibration, return the meter to a
factory Service/Sales Support Branch or authorized service
station.
8

ENGLISH

Accuracy
±0.2% ± 10 dgts
±0.2% ± 5 dgts
±1.0% ± 5 dgts
6 mm
‑10°C ‑ 50°C
‑40°C ‑ 60°C
0.1 x specified accuracy /°C (<18°C or >28°C)
0‑2000 m
1 m
2 AA
approx. 25 h
360 g
This appliance is not intended for use by persons (including
children) with limited physical, sensory or mental capabilities or
insufficient experience and/or knowledge unless they are
supervised by a person who is responsible for their safety or
have been instructed by them in the safe use of the tool.
Children should be supervised in order to ensure that they do
not play with the tool.
SAFETY INSTRUCTIONS Battery
This tool is designed to be powered by 2 AA batteries properly
inserted into the instrument. Do not attempt to use with any
other voltage or power supply.
Do not leave batteries within the reach of children.
Do not mix new and used batteries. Do not mix brands (or
types within brands) of batteries.
Do not mix rechargeable and non‑rechargeable batteries.
Install batteries according to polarity (+ / –) diagrams.
Properly dispose of used batteries immediately.
Battery acid may leak from damaged batteries under extreme
load or extreme temperatures. In case of contact with battery
acid wash it off immediately with soap and water. In case of eye
contact rinse thoroughly for at least 10 minutes and
immediately seek medical attention.
Do not try to replace the batteries if the surface of the
instrument is wet.
Disconnect all the cords and cables from the object under test
and power off the instrument before opening the battery cover
for battery replacement.
Do not incinerate or dismantle batteries.
SPECIFIED CONDITIONS OF USE
The Measuring Instrument can be used for following
measurements: Current DC. This instrument may only use for
measuring the previously indicated measurements.
DISPLAY
Battery life indicator
4 to 20 mA Span Percentage
Primary measurement
Units indicator
Hold indicator
BUTTONS
Power ON/OFF
Zero Button
H O L D
Z E R O
Hold button
ZERO Button
Always press the ZERO button before clamping around a
conductor. Pressing the ZERO button removes the residual DC
current offset to ensure a more accurate reading.
HOLD Button
Press the HOLD button to freeze the reading. The reading will
be held regardless of subsequent variation in input. HOLD is
displayed with the reading. To exit Hold mode, press the HOLD
button again.
Sleep Mode
The display will turn off about 5 minutes after the last button
operation to conserve battery life. Press any button to turn the
display back on.
Auto Power Off
The meter will completely power off about 15 minutes after the
last button operation to conserve battery life.
4 to 20 mA Span Percentage
22 mA .......110%
20 mA ...... 100%
16 mA ........ 75%
12 mA ........ 50%
8 mA ........ 25%
4 mA .......... 0%
0 mA ....... -25%
MEASURING
1. Remove the clamp from the clamp storage pocket.
2. Press the
button to turn on the meter. Always press the
button before placing the clamp around a conductor to
H O L D
Z E R O
H O L D
Z E R O
take a new measurement.
3. Keep fingers behind tactile barrier.
4. Open the jaws and clamp them around the conductor under
test.
NOTE: Do not clamp over more than one wire at a time.
Irregular results will occur.
5. The reading is displayed. A negative number indicates
current flow in the opposite direction of the arrows on the
clamp.
6. Press the
button to turn off the meter. Remove the jaws
from the conductor.
H O L D
Z E R O
CAUTION! Maximum conductor size is 6 mm diameter.
During measurement, keep the jaws fully closed to ensure
accurate measurements.
MAINTENANCE
Use only Milwaukee accessories and Milwaukee spare parts.
Should components need to be replaced which have not been
described, please contact one of our Milwaukee service agents
(see our list of guarantee/service addresses).
If needed, an exploded view of the tool can be ordered. Please
state the Article No. as well as the machine type printed on the
label and order the drawing at your local service agents or
directly at: Milwaukee Electric Tool, Max‑Eyth‑Straße 10,
D‑71364 Winnenden, Germany.
SYMBOLS
Please read the instructions carefully before
starting the machine.
Do not dispose of electric tools together with
household waste material! In observance of
European Directive 2002/96/EC on waste
electrical and electronic equipment and its
implementation in accordance with national law,
electric tools that have reached the end of their
life must be collected separately and returned to
an environmentally compatible recycling facility.
WARNING
To avoid electrical shock remove test leads
before opening case or removing battery pack/
battery.
Ground
Double isolated
Cat II
Classification of transient over‑ voltages, based
on nominal line voltage to earth.
Do not apply around or remove from
HAZARDOUS LIVE conductors.
Direct Current
ENGLISH
9