NOTE
The connector B is a push-pull connector. Align the red dot on top of the male connector to the
upright position to fit the guide on the female connector. Do not screw into the connector. To
disconnect, pull out using the metal end of the cable. Do not twist in any direction.
2. Attach the WaveOne
contra-angle operating instructions).
3. Disinfect the keypad and micromotor before first use and before each use on a new patient
(for details see chapter 6.9).
6.3.3. Battery
The WaveOne
endo motor is powered by a rechargeable Nickel Metal Hydride (NiMH)
™
battery.
WARNINGS
Do not open the device to replace the battery for any reason, as there is the risk of a short-
circuit. Opening the device will invalidate the warranty.
The battery can be replaced only by an authorized service center.
NOTE
Fully charge the battery prior to first use
When connecting for charging, always check that the green LED on the battery charger
is ON to show that the battery charger is properly connected to the power supply.
The battery LED appears in 3 colors indicating the current battery condition:
Green: Indicates a battery capacity of 20% - 100 %
Flashing red: Battery needs to be charged. Battery power will run out in a few minutes. Before
the motor switches off, there is an audible warning signal and the display will read:
(MAILLEFER 6:1) contra-angle to the micromotor (see separate
™