Note: All parameters needed to amend will use the similar way to achieve.
All chapters about setting manual temperature here after will be operated according to this.

4.4.5 System Setup

This function is GLP regular setting includes setting electrode calibration interval, Balance
Reading measuring condition, operator's number, whether to delete stored data automatically and
whether open GLP function.
Press 'SETUP' key, select 'System Setup' item, press 'Enter' to enter into system setting mode. It
is displayed as follows:
Press relevant item to amend corresponding parameter. When amendment finishes, press 'X' (quit)
to exit, then return to start mode.
Electrode calibration interval means meter will remind user the time interval of calibrating
electrode. Meter will automatically calculate the time from previous calibration till now. In case
previous calibration time exceeds the time interval set by user, meter will jump a reminding
window to remind user take care and re-calibrate electrode. The unit of electrode calibration
interval is hour. The displayed pH electrode calibration interval is 4 hours.
The operator's number is a three digit number, the range is between 000 ~ 200, all operation
records contain operator's number.
Automatic delete stored data function means when stored data reaches the amount that the meter
Display chart of 'Set Manual Temperature'
Display chart of 'System Setup'
67