In 'Setup Measuring Mode', select dissolved oxygen parameter such as dissolved oxygen,
saturation etc... to start measuring of DO which is displayed as follows:
On upper side of the screen is present measuring data, system time, DO measuring result and
relevant current DO value, current temperature value. On right side is press key including 'VIEW',
'STD', 'STORE', 'PRINT' and 'SETUP'.
Press 'VIEW' to view conductivity measuring parameter, set current air pressure and salinity
etc... (Please refer to 'View solution's Measuring Parameter' for details).
Press 'STD' to re-calibrate DO's zero oxygen, full scale etc... (Please refer to 'Dissolved
Electrode Calibration' for details).
During measurement you can press 'SETUP' to set manual temperature.
If you select Continue Mode (Please refer to 'Setup Measuring Mode' for details), When
measuring result stable, you can press 'STORE' to save current measuring result. If meter is
connected with TP-16 printer before turned on, you can press 'PRINT' key, meter will print
present measuring data.
If you select Timed Reading Measuring Mode (Please refer to 'Setup Measuring Mode' for
details), meter will automatically measure, calculate and display measuring result. When the time
interval meets the time set by user, meter will automatically store measuring result. If meter is
connected with TP-16 printer before turning on, meter will automatically print measuring data and
start next measuring when printing finishes.
If you select 'Balance Reading measuring mode' (Please refer to 'Setup Measuring Mode' for
details), When measuring starts, meter will automatically measure, calculate and display
measuring result. In case the measuring suits the set balance condition, the measuring finishes! At
this time, you can press 'STORE' to save current measuring result. If meter is connected with
TP-16 printer before turning on, you can press 'PRINT' key, meter will print present measuring
data.
Press 'X' on top right corner and the press 'ENTER' to finish the measuring.