Sirona Dental Systems GmbH
Supplement to Operating Instructions
empty
62 38 344 D3439
D3439.201.04.03.02 04.2010
2
Using the manual block clamp
Store the block clamp tool in the corresponding holder.
You can attach the holder to a suitable location with the adhesive pad. Clean
and degrease the contact surface beforehand.
NOTICE
Wear of the ball pressure screw
The high clamping forces cause wear of the ball pressure screw.
➢ Replace the ball pressure screw every 500 clamping operations.
Blocks with 10 mm diameter block holders
Fastening the block
NOTICE
Fasten the block tightly
If the block is not tightened sufficiently, this may result in falsification of the
milling result and fracturing of its ceramic material.
➢ Fasten the block tightly using the block clamp tool with torque wrench un -
til you hear a click.
➢ Check to make sure that the block is seated very firmly.
1.
Place the block (A) directly into the block clamp.
Instructions for 10 mm diameter
2.
Clamp the ceramic block securely with the ball pressure screw (B). Use
the block clamp tool with torque wrench for this purpose.
The block is pressed laterally against the contact surface of the block
clamp and simultaneously pulled in axially. The plate of the block hol -
der thus rests on the block clamp.
Using the manual block clamp
3