Vacuum System; Care; Disinfection When Using Dry Aspiration - Sirona C1+ Manual De Instrucciones

Ocultar thumbs Ver también para C1+:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

3 Vacuum system

3.1 Care

3
Vacuum system
3.1
Care
3.2

Disinfection when using dry aspiration

8
A
8
After each patient
To ensure that the vacuum system is always ready for
use, a large glass of cold, clear water must be aspirated
through the used suction hoses after each patient,
especially after each intervention during which the
patient loses blood and after using hydrogen peroxide
(H
O
).
2
2
In long-lasting treatments a glass of water must be
aspirated at least every 60 min.
Daily before the mid-day break and
after the practice has closed
The vacuum system is subjected daily to germ-carrying
secretions, saliva and blood.
For hygienic reasons the system must therefore be
disinfected with a detergent/disinfectant approved by
Sirona.
ATTENTION
Use only cleaning and disinfecting agents which have
been approved by Sirona !
Do not use any detergents or other disinfectants!
• Prepare the cleaning solution in a separate container
according to the manufacturer's instructions and mix
thoroughly.
• Remove the gold trap.
• Clean the cuspidor.
• Plug in the cleaning adapter up to the stop. For the sa-
liva ejector hose, insert the intermediate piece (8) in
the cleaning adapter.
• Fill the cleaning solution up to the filling limit (A) in the
cuspidor (approx. 0.75 liters)
• Remove the suction hoses from their receptacles and
attach them to the side of the cleaning adapter, if pos-
sible simultaneously.
• The cleaning solution is aspirated.
The cleaning process takes several minutes.
Sirona Dental Systems GmbH
.
59 06 511 D 3386
D 3386.103.01.06.02

Publicidad

Tabla de contenido
loading

Tabla de contenido