10.3 Use of foot switches

For safe operation of the foot switches, observe the following:
Ensure that only Class AP foot switches are used in the environment described as „medical".
(1) Medical environment
Medical environment 1 includes the area where flammable mixtures with air can cause an explosion
with a normal ignition energy. The anaesthetic gases and the vapours of the disinfectants are heavier
than the air and fall to the floor.
There is no risk of explosion outside this area because of the stronger dilution of these substances.
11

Review of the technical status

Before any procedure, check the cables, accessories (handles, electrodes, forceps) and foot switches
for any signs of visible irregularity. Do not use cables or instruments with broken or defective insulation.
These can result in burns if they come into contact with the skin!

11.1 Mechanical faults

If damage occurs to switches, connecting sockets, houses or the touchscreen, or if the unit is dropped
on the floor, please contact your authorized dealer for advice on further use.
The dealer can arrange for a detailed technical check to be performed
The manufacturer accepts no liability for calibration work or repairs to the unit carried out by the user.