2. Insert the VisiClear™ Filter into filter receptacle. Be sure that the filter is seated completely
against the bottom of the filter chamber and clip is fully engaged.
WARNING: This device is not intended for evacuation of fluid. If fluid is expected to be
aspirated using the VisiClear™ Filter or the BUFFALO FILTER® VisiClear™ system, fluid
collection devices must be installed with the vacuum hose assembly. Failure to install a fluid
collection device may cause filter blockage and/or electrical damage.
Filter Removal Instructions:
1. After the VisiClear™ Filter has been exhausted and requires changing, turn the smoke
evacuation system off and disconnect any accessory tubing attached to the filter.
2. Depress the tab and pull the VisiClear™ Filter from the smoke evacuation system and
dispose of in accordance with hospital policy. The VisiClear™ Filter may be disposed of or
incinerated.
3. Clean the unit with appropriate germicide prior to re-use. Follow the indicated instructions
for maintenance and installation of a new VisiClear™ Filter.
CAUTION: Using any other filter or accessory not supplied by BUFFALO FILTER® may
cause damage to the system and/or cause the system to be inoperable and may void the
warranty.
WARNING: The VisiClear™ Filter should be changed when the Filter Life Indicator touch
screen indicates an expired filter. Failure to change this filter may result in decreased
efficiency and contamination of the electric motor, vacuum pump, and sound absorbing
media within the system, or non-operation of smoke evacuator.
Buffalo Filter® VisiClear™ Operator's Manual