Standard Accessories:
• Operator's Manual
• Power Cord
• Pneumatic Footswitch
Please contact BUFFALO FILTER® Customer Service to purchase the following accessories:
• Replacement Filters
• EZLink™ Automatic Activation Device
• Hoses, Tubing, Laparoscopic Kits, Adapters, Wands & Other Accessories

1.3 Operational Information

The operational information contained in this section is intended for the customer review of
regulatory issues. The information pertains to the use of the products both domestically and
internationally:
1. The BUFFALO FILTER® VisiClear™ Surgical Smoke Plume Evacuator(s) complies
with IEC60601-1:2005 electrical specifications in the following systems:
100/120 VAC, 50/60 Hz, 220/240 VAC, 50/60 Hz
2. Type of protection against electrical shock: Class I
3. Degree of protection against electric shock: Type CF Applied Part
4. Degree of protection against ingress of water: IPX1
5. Method of sterilization or disinfection recommended by BUFFALO FILTER®:
Unplug Unit. Wipe unit with a damp cloth containing mild disinfectant solution
or soapy water. Wipe dry with a clean cloth. Do not steam sterilize.
6. Degree of safety of application in the presence of flammable anesthetic mixture with air or
with oxygen or nitrous oxide: Not Suitable
7. Mode of operation: Continuous
8. The fuses on the circuit board are (F1, F20 to be serviced by an authorized BUFFALO
FILTER® technician as follows:
100/120 VAC, 50/60 Hz use 10 Amp 250 Volt Fuse (Slo-Blo)
with Interruption Rating: 1500A@250V
220/240 VAC, 50/60 Hz use 8 Amp 250 Volt Fuse (Slo-Blo)
with Interruption Rating: 1500A@250V
9. The fuse on the transformer circuit (F6) to be serviced by an authorized BUFFALO FILTER®
technician as follows:
100/120 VAC, 50/60 Hz use 100 Amp 250 Volt Fuse (Fast-Act)
with interruption rating: 1500A@250V
220/240 VAC, 50/60 Hz use 50Amp 250 Volt Fuse (Fast-Act)
with interruption rating: 1500A@250V
10. The fuses on the motor circuit are to be serviced by an authorized BUFFALO FILTER®
technician as follows: 220/240 VAC, 50/60 Hz use 3,15 Amp 250 Volt Fuse (Fast-Acting), (F3)
with Interruption Rating: 1500A@250V
11. This equipment needs special precautions regarding Electro Magnetic Compatibility and
needs to be installed according to EMC information found in this manual.