Corrective Action - Buffalo filter VisiClear Manual Del Operador

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 62
3.4 Troubleshooting the System – see below.
PROBLEM
1. Smoke Evacuation
System is ON but
suction is minimal
or none.
2. Smoke Evacuation
System does not
function even though
suction ON button is
depressed.
Buffalo Filter® VisiClear™ Operator's Manual
POTENTIAL CAUSE
1. Filter is not seated
completely.
2. Filter is clogged.
3. Vacuum hose or tube is
clogged.
4. Motor/blower is
obstructed.
1. Not plugged into an
electrical outlet.
2. Fuses are blown.
3. Electronic system
failure.
4. Filter life has expired or
invalid filter inserted.

CORRECTIVE ACTION

1. Re-install VisiClear™ Filter, press firmly into place
and fully engage clip.
2. Replace filter with a genuine Buffalo Filter
VisiClear™ Filter.
3. Replace vacuum hose or tube with genuine
Buffalo Filter products.
4. Call BioMed or BUFFALO FILTER® Technical
Services at 1.800.343.2324 or 716.835.7000.
1. Check power outlet and connection to rear or side
panel of the machine.
2./3. Call BioMed or BUFFALO FILTER® Technical
Services at 1.800.343.2324 or 716.835.7000.
4. Replace filter with a genuine Buffalo Filter
VisiClear™ Filter.
27