Technical characteristics (cont'd)
X-Y mode
Phase error
Range of sampling rates in
Equipment frequency counter
Reading resolution
Range of operating
frequencies of the meter
Type of signal
Control panel function
Backup/Restoration
Measurement system
Automatic measurements
Measurements by cursor
Two-channel digital Oscilloscopes
X-Y inputs
CH1 (X)/CH2 (Y)
±3 degrees
XY mode sampling rate:
XY mode
12,5kSa/s ~ 250MSa/s (MemDepth: Normal)
500,5kSa/s ~ 250MSa/s (MemDepth: LongMem)
6 Bytes
Accuracy
±0.01%
With DC coupling, the range of operating frequencies of the meter is
from 10Hz to the max frequency of the bandwidth.
All triggering source signals that generate triggering events (except:
"pulse width" trigger and "video" trigger).
Auto Set
Self-adjustment of the Vertical and Horizontal systems and of the
trigger
Internal backup capacity: 2 REFerences, 20 configurations, 20
Traces. Possibility of saving traces and configurations to external
USB flash memory.
Vpp, Vmax, Vmin, Vamp, Vtop, Vbase, Vavg, Mean, Crms, Vrms,
(32 types)
ROVShoot, FOVShoot, RPREShoot, FPREShoot, Rise time, Fall time,
Freq, Period, +Wid, -Wid, +Dut, -Dut, BWid, Phase, FRR, FRF, FFR,
FFF, LRR, LRF, LFR, LFF
Modes: Manual, Track, and Auto
27