Troubleshooting; Isa100 Wireless ™ Standard Objects; Measured Value Status (Pv.status) - Dräger Polytron 6100 EC WL Instrucciones De Uso

Ocultar thumbs Ver también para Polytron 6100 EC WL:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

11
Display
Access
PV
READ
PV
READ
PV
READ
PV
READ
PV
READ
READ
DIAG_STATUS
READ
DIAG_STATUS
READ
DIAG_STATUS
READ

11.2 Measured value status (PV.STATUS)

Meaning "Quality" (bit 6 and 7):
0: Bad - No transfer of the measured value.
1: Uncertain - Measurement quality lower than normal.
2: Good - Measurement quality good, but alarm condition could be present.
Quality
Substatus
Bits 6 and 7 Bits 2 to 4
0
0
0
1
0
2
0
3
0
4
0
5
0
6
Dräger Polytron 6100 EC WL, Dräger Polytron Repeater ISA
standard objects
Byte no.
1
2
3
4
5
1
2
3
4
PV
Cause
NaN
Start sequence active.
Start sequence failed.
NaN
Sensor lock feature is enabled. A
sensor with a different part number
was inserted.
NaN
- (Bit used by the gateway)
NaN
Serious device fault, various
causes.
NaN
Sensor not detected or defective.
NaN
Communication with gas detector
interrupted, last usable measured
value available (used by the gate-
way)
NaN
Communication with gas detector
interrupted, last usable measured
value not available (used by the
gateway)
Content
Modbus input
register
PV_Value (1/4)
Register N+1
PV_Value (2/4)
PV_Value (3/4)
Register N+2
PV_Value (4/4)
Register M
Diag_Status (4/4)
Remedy
Wait for start sequence.
A Restart the gas detector.
B Have the gas detector checked by Dräger-
Service.
Use a sensor with the same part number or dis-
able the sensor lock feature.
-
A Restart the gas detector.
B Reset the gas detector to the factory set-
tings.
C Reset the sensor to the factory settings.
D Update firmware.
E Have the gas detector checked by Dräger-
Service.
A Check the sensor connection.
B Remove and reinstall the sensor.
C Change the sensor.
A Check network configuration.
B Check positioning of gas detector antenna
and gateway.
C Restart the gas detector.
A Check network configuration.
B Check positioning of gas detector antenna
and gateway.
C Restart the gas detector.

Troubleshooting

Data type
UINT (16 bit)
32-bit float (4 bytes), Float32
acc. toIEEE 754
UINT (16 bit)
UINT (16 bit)
en

Publicidad

Capítulos

Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

Polytron repeater isa

Tabla de contenido