12.3 RF (Radio Frequency) Interference
EMC disturbance can affect the performance of the ARTROMOT®-S3. To prevent the user
from unacceptable risks the device runs performance checks regularly during operation.
This is what happens, if a problem is identified:
− an audio signal sounds
− the unit stops immediately
− the message ERROR and an error code appear on the display
In this case try to quit the error with the STOP key. If the error persists you may attempt to
restart the unit by turning it briefly off and on with the power switches.
WARNING!
m
Portable RF communications equipment (including peripherals such as antenna cables
and external antennas) should be used no closer than 30 cm (12 inches) to any part of the
ARTROMOT®-S3, including cables specified by the manufacturer. Otherwise,
degradation of the performance of this equipment could result.
Notice!
(for home use)
Portable and mobile RF communication systems may affect the ARTROMOT®‑S3 device.
For this reason, make sure that wireless communication equipment, such as wireless home
network devices, cell phones, cordless telephones and their base stations and walkie‑talkies
operate at a minimum distance of 3.3 m from the device. (Calculated on the basis of the
maximum power output of a typical cell phone of 2 W).