Protocol Details - Comfort Model - DJO ARTROMOT-S3 Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 126
LEVEL 5:
– Oscillation
– Stretching in abduction
– Stretching in
internal rotation
– Stretching in
external rotation
LEVEL 6:
– Transport setting
– Synchronized/
non‑synchronized mode
– Total therapy time
– Service menu
5.5 Protocol details –
Comfort model
• To access the programming level you
need, press the MENU key repeatedly.
• The treatment parameters are selected
with the different parameter keys.
• To change the treatment values, press
the + / – keys.
• To enable/disable a function, press the
respective parameter key again.
• To save the entered values, press the
STOP key.
All special functions are disabled in the
standard models.
Note!
LEVEL 1: equivalent to level 1
of the Standard model (see: 5.3)
LEVEL 2: equivalent to level 2
of the Standard model (see: 5.3)
LEVEL 3: equivalent to level 3
of the Standard model (see: 5.3)
LEVEL 6: equivalent to level 4
of the Standard model (see: 5.3)
LEVEL 4:
n Warm-up protocol
During warm‑up, the patient will slowly
become used to the set maximum
values for adduction/abduction and
internal/external rotation.
After activation of the warm‑up pro‑
tocol the carriage moves between the
maximum adduction and abduction
positions and the center position of the
programmed range of motion (adduc‑
tion/internal rotation and abduction/
external rotation).
With each cycle, 3° is added to the
abduction/external rotation range of
motion until the maximum abduction/
external rotation positions are reached.
The adduction and internal rotation
ranges are completed in each cycle.
Once the maximum abduction and
external rotation values have been
attained, the carriage switches to the
normal mode.
If a therapy time is activated, the system
restarts with the warm‑up protocol
when the therapy time has elapsed and
the devices is started again.
During warm‑up, the display will show
the symbol
hand corner.
Default: disabled
n Isolation protocol
For this special function, both motors
are turned on, but they will never move
simultaneously.
These are the steps of the special
function:
• At first motor A moves to the pro‑
gram med maximum values for ad‑
duction and abduction for ten cycles,
then it stops. In this phase, motor B is
disabled (symbol for motor B: OFF)
• Then motor B will move to the pro‑
grammed maximum values for inter‑
nal and external rotation for 10 cycles
and stops, after this the full cycle will
start over with motor A. Motor A is
disabled while motor B is operating
(symbol for motor A: OFF)
65
in the upper, left‑

Publicidad

Tabla de contenido
loading