Technical inspections
Due to their design, the bearings and
joints of the ARTROMOT®‑S3 do not
require maintenance and all materials are
protected from corrosion, BUT: For safety,
the devices require regular maintenance.
To maintain the functional and operational
safety, check all components for damage
and loose connections at least once a year.
These checks should be performed by
persons with adequate training and
experience. Damaged and worn parts must
immediately be replaced with original
spare parts by authorized, specially trained
staff.
Warning!
m
Patient hazard, equipment malfunction
and damage
− Refer repair and maintenance to
authorized persons. Persons are
authorized after training by a special‑
ist trained and commissioned by the
manufacturer.
ORMED GmbH will make all docu‑
ments required for servicing, such
as circuit diagrams, parts lists, de‑
scriptions or calibration instructions,
available to authorized experts.
These checks should be performed by
persons with adequate training and
experience. Damaged and worn parts
must immediately be replaced with
original spare parts by authorized staff.
Other than that, the manufacturer
does not require any other regular
maintenance.
Note!
Regarding further technical or other
inspections or inspection intervals,
observe applicable local requirements,
such as IEC 62353, DGUV3 or similar
regulations to be observed by users of
medical equipment or electric devices.

Fuse replacement

Warning!
m
Patient hazard, equipment malfunction
and damage
The replacement of fuses must be
referred to specialists as defined in
IEC 60364 or other applicable standards
(e.g. biomedical technicians, electricians,
electronics installers).
Fuses used must be
S# < 20.000: T1A L250Vac
S# > 20.000: T2A H250Vac
fuses.
• Before replacing fuses, turn off the
ARTROMOT®‑S3 and disconnect the
device from the power line.
• Use an appropriate tool to remove the
fuse holder (8) situated between the
power switch (7) and the power connec‑
tor (6) (Fig. 1).
• Replace the fuses and reinsert the fuse
holder (Fig. 2). Check that the fuser holder
locks properly into place.