1.3 EN/IEC 60601-1-2:2007 Sub-clause 5.2.2.2 Table 4:
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The A
S
Temperature Management System unit is intended for use in the
®
rctic
un
electromagnetic environment specified below. The customer or the end user of the
A
S
Temperature Management System unit should assure it is used in such an
®
rctic
un
environment.
Immunity
IEC60601
Compliance
Test
test level
Level
Conducted
3Vrms
3Vrms
RF
150kHz to
150kHz to
IEC 61000-
80MHz
80MHz
4-6
3V/m
3V/m
Radiated
80MHz to
80MHz to
RF
2.5GHz
2.5GHz
IEC 61000-
4-3
NOTE 1 At 80MHz and 800MHz, the higher frequency range applies
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from objects, structures and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/
a
cordless) telephones and land mobile radios, amateur radio, AM and FM radio
broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess
the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the
A
S
Temperature Management System unit is used exceeds the applicable
®
rctic
un
RF compliance level above, the A
should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as re-orienting or relocating the
A
S
Temperature Management System unit.
®
rctic
un
Over the frequency range 150kHz to 80MHz, field strengths should be less than 3V/m.
b
12
Intended Electromagnetic Environment
Portable and mobile RF communications
equipment should be used no closer to
any part of the A
S
®
rctic
un
Management System unit, including cables,
than the recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance
d = 1.2√P
d = 1.2√P
80MHz to 800 MHz
d = 2.3√P
800MHz to 2.5GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended minimum
separation distance in meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey
, should be less than the compliance
a
level in each frequency range.
Interference may occur in the vicinity
of equipment marked with the following
symbol:
S
Temperature Management System unit
®
rctic
un
1.4 EN/IEC 60601-1-2:2007 Sub-clause 5.2.2.2 Table 6:
Recommended separation distances between portable and mobile RF
communications equipment and the A
System unit
RF communications equipment can effect medical electrical equipment. The
A
S
Temperature Management System unit is intended for use in an
®
rctic
un
electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the A
can help prevent electromagnetic interference by maintaining a minimum distance
between the portable and mobile RF communications equipment (transmitters) and
the A
S
rctic
un
according to the maximum output power of the communications equipment.
Temperature
Rated maximum
output power of
transmitter in watts
(W)
0.01
0.1
1.0
10
100
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in meters (m) can be estimated using the
b
equation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects
and people.
O P E R A T O R ' S M A N U A L
S
Temperature Management
®
rctic
un
S
Temperature Management System unit
®
rctic
un
Temperature Management System unit as recommended below,
®
Separation distance according to frequency of transmitter in
meters (m)
150kHz to 80MHz
80MHz to 800MHz
d = 1.2√P
d = 1.2√P
0.12
0.12
0.38
0.38
1.2
1.2
3.8
3.8
12
12
800MHz to
2.5GHz
d = 2.3√P
0.23
0.73
2.3
7.3
23