Main Safety Alarms

While there are multiple alarms and safety features in the A
Temperature Management System, there are five main safety alarms that
will place the device into Stop mode until the condition is addressed.
Alarm
Specification
High Patient Temperature
39.5°C (103.1°F)
Low Patient Temperature
31.0°C (87.8°F)
High Water Temperature
42.5°C / 44°C (108.5°F / 111.2°F)
Low Water Temperature
3.0°C / 3.5°C (37.4°F / 38.3°F)
System Self-Test Failure
At device power ON
Each time the A
S
Temperature Management System is powered
®
rctic
un
On, a system self test for the independent safety alarm is automatically run.
This test simulates a "water high temperature" fault situation on both the
primary and secondary water temperature sensors. Both the primary
and secondary safety systems must respond to the fault and be
verified by the opposing safety system. If either safety systems do not
respond appropriately either an alarm 80 or 81 will be issued. Contact
Customer Support.

Non-Recoverable Alarms

If an Alarm condition occurs that prevents proper use of the device or proper
patient treatment (such as the five main safety alarms discussed above),
the system is placed into Stop mode and will not allow therapy to continue.
This type of Alarm is known as Non-Recoverable. If this situation occurs,
cycle the device power (turn device Off then On). If the alarm recurs contact
Customer Support.

Recoverable Alarms

Other Alarms that temporarily Stop the device until the user is able to
correct the cause and clear the Alarm are classified as Recoverable. If the
condition that initiated the alarm is not addressed and problem persists, the
Alarm will recur.
If a Recoverable Alarm occurs:
1) When an alarm is issued the device is placed into Stop mode.
2) Read the displayed instructions.
3) Note the Alarm number.
4) Press the Close button to clear the alarm.
5) Follow the instructions to correct the alarm condition. Perform the
actions in the order listed until the alarm condition is resolved.
6) Once you have cleared the alarm, press the Start button in the therapy
window to restart therapy. You will hear a tone and a voice stating
"Therapy Started". Additionally, the active therapy window and the
A
S
Temperature Management System icon will blink.
®
rctic
un
7) If the condition does not resolve, contact Customer Support.

Alerts

Alerts are denoted by an audio signal that repeats every 25 seconds.
The Alert screen will appear that displays the alert number, alert title,
a description of the problem that triggered the alert, and solutions and
instructions for troubleshooting and resolving the alert condition.
If an Alert occurs:
1) Read the displayed instructions.
2) Note the Alert number.
3) Press the Close button to clear the alert.
4) Follow the instructions to correct the alert condition. Perform the actions
in the order listed until the alarm condition is resolved. If the condition
does not resolve, contact Customer Support.
5) Refer to the A
S
Temperature Management System Help
®
rctic
un
screens for additional information regarding alarms and alerts.
Chapter 7 – Maintenance and Service
S
®
rctic
un

Cleaning and Maintenance

Routine cleaning and preventive maintenance should be performed on the
A
S
rctic
un
at a minimum. This consists of cleaning the external surfaces, accessories
and chiller condenser, inspecting the device, and replenishing the internal
cleaning solution that suppresses microorganism growth in the water reservoir
and hydraulic circuit. See the A
Service Manual for additional information.
External Surfaces
Clean the exterior body of the control module, fluid delivery lines, power
cords and temperature cables using a soft cloth and mild detergent or
disinfectant according to hospital protocol.
Condenser
A dirty chiller condenser will significantly reduce the cooling capacity of the
control module.
To clean the condenser, wipe the dust from the exterior grill using a soft
cloth. Depending on the quality of your institution's air, periodically remove
the back cover and vacuum or brush the condenser fins. At a minimum the
condenser fins should be cleaned annually. Maintenance activities should be
performed by qualified personnel.
Device Inspection
Periodically inspect the external areas of the device for damaged, loose
or missing parts, and frayed or twisted power cords and cables.
Discontinue using the device displaying one or more of the above
conditions until the problem is corrected and has been verified to be
operating correctly.
Replenish Internal Cleaning Solution
Contact Customer Service to order internal cleaning solution.
To replenish the internal cleaning solution:
1) Drain the reservoir.
Turn control module power Off.
Attach the drain line to the two drain ports on the back of the control
module. Place the end of the drain line into a container. The water will
passively drain into the container.
2) Refill the reservoir.
From the Hypothermia therapy screen or the Normothermia therapy
screen, press the Fill Reservoir button.
The Fill Reservoir screen will appear. Follow the directions on
the screen.
Add one vial of A
solution to the first bottle of sterile water.
The filling process will automatically stop when the reservoir is full.
Continue to replace the bottles of sterile water until the filling
process stops.
When the Fill Reservoir process is complete, the screen will close.

Preventative Maintenance

Use of the A
2,000 hours, without conducting preventative maintenance, may result in
failure of certain system components and failure of the system to function
as intended. To maintain system performance, the A
Management System requires periodic service of key components.
For additional information, please refer to http://www.medivance.com/manuals,
call 1-800-526-4455 or contact your local B
Temperature Management System control module every 6 months
®
S
Temperature Management System
®
rctic
un
S
Temperature Management System cleaning
®
rctic
un
S
Temperature Management System in excess of
®
rctic
un
E N G L I S H
S
Temperature
®
rctic
un
representative.
®