Technical Information
This vent valve waits before opening a minimum time of about 5 sec.
This time can be increased up to about 220 min. by means of a setting of the
controller software (optional). To install the vent valve, unscrew the threaded plug
(see figure below).
Figure 53
Then screw the vent valve into the pump and tighten it using a 16 mm hexagonal
spanner with a torque of 2.5 Nm.
Figure 54
The vent kit shown in the picture is the 9699834
Do not overtighten the valve as this may damage the thread on the pump.
CAUTION
Then connect the cable from the valve to the suitable connector on the controller.