Hillrom Welch Allyn KleenSpec 800 Serie Manual Del Usuario página 17

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 77
Instructions for use
Power frequency
(50/60 Hz) magnetic
field IEC 61000-4-8
Note: U
is the a.c. mains voltage prior to application of the test level.
T
The 800 Series is intended for use in the electromagnetic environment specified below. The customer or the
user of the 800 Series should assure that it is used in such an environment.
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Electromagnetic immunity
70 % U
; 25/30 cycles
70 % U
T
Single phase: at 0°
0 % U
; 250/300
0 %U
; 250/300 cycle
T
T
cycle
30 A/m
30 A/m
Electromagnetic immunity
IEC 60601 test level Compliance level
3 Vrms
3 Vrms
150 kHz to 80 MHz
6Vrms in ISM and
6Vrms .
amateur radio bands
between 150 kHz
and 80 MHz.
10 V/M, 80 MHz to
10 V/M
2.7 GHz
; 25/30 cycles
T
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or hospital
environment.
Electromagnetic environment -
guidance
Portable and mobile RF communications
equipment should be used no closer to
any part of the 800 Series, including cables,
than the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
800 MHz to 2.7 GHz
80 MHz to 800 MHz
where P is the maximum output power
rating of the transmitter in watts (W) and d
is the recommended separation distance
in meters (m). Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey
less than the compliance level in each
frequency range
English 13
a
, should be
b
. Interference may occur

Publicidad

Tabla de contenido
loading

Tabla de contenido