12 English
IEC 61000-3-3
a
The 800 Series contains a 5-GHz orthogonal frequency-division multiplexing transmitter or a 2.4-GHz
frequency hopping spread-spectrum transmitter for the purpose of wireless communication. The radio is
operated according to the requirements of various agencies, including FCC 47 CFR 15.247 and Radio
Equipment Directive 2014/53/EU. The transmitter is excluded from the EMC requirements of 60601-1-2, but
should be considered when addressing possible interference issues between this and other devices.
The 800 Series is intended for use in the electromagnetic environment specified below. The customer or the
user of the 800 Series should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
Welch Allyn® KleenSpec 800 Series Cordless Illumination System
Electromagnetic emissions
Electromagnetic immunity
IEC 60601 test level Compliance level
±8 kV contact
±8 kV
±15 kV air
±15 kV
±2 kV for power
±2 kV
supply lines
±1 kV for input/
±1 kV
output lines
±0.5 kV, ±1 kV
±1 kV
Line- to -line
±0.5 kV, ±1 kV, ±2 kV
±2 kV
Line-to-ground
0 % U
; 0.5 cycle
0 % U
; 0.5 cycle
T
T
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0 % U
; 1 cycle
0 % U
; 1 cycle
T
T
WARNING This equipment/system is
intended for use by healthcare professionals
only. This equipment/ system may cause radio
interference or may disrupt the operation of
a
nearby equipment
. It may be necessary to
take mitigation measures, such as re-orienting
or relocating the 800 Series or shielding the
location.
Electromagnetic environment -
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the 800 Series
requires continued operation during
power mains interruptions, it is
recommended that the 800 Series be be
powered from an uninterruptible power
supply or a battery.